不同厚度全锆牙合贴面的断裂载荷及失效模式研究

曾晓燕1,施佳丽1,王柏华1,李梦圆1,游苒2,胡建1,李林1   

  1. 1. 南京医科大学附属口腔医院
    2. 南京医科大学附属医院
  • 收稿日期:2018-07-04 修回日期:2018-10-26 出版日期:2018-12-25 发布日期:2019-01-07
  • 通讯作者: 胡建 E-mail:hujianmjc@sohu.com
  • 基金资助:
    江苏省高校优势学科建设工程资助项目;江苏省高校大学生创新创业训练计划项目

Effects of thickness on fracture strength and failure mode of full-contour zirconia occlusal veneers

  • Received:2018-07-04 Revised:2018-10-26 Online:2018-12-25 Published:2019-01-07
  • Contact: Jian Hu E-mail:hujianmjc@sohu.com

摘要: 探究贴面厚度以及牙体预备形式对全锆贴面的抗折性能及失效模式的影响。方法:预备标准人工磨牙形成3组不同形式的预备体:A组面均匀磨除0.5 mm,B组面均匀磨除0.5 mm,轴面包绕高度1.5 mm,C组面均匀磨除0.5 mm,并预备髓腔固位。采用CAD/CAM系统,通过Dental Designer软件,将每组分别设计为0.5 mm、1.0 mm、1.5 mm三种不同厚度,共制作出72个贴面;测试不同牙体预备形式及不同厚度的贴面的断裂载荷,并分析其失效模式。结果:不同牙体预备形式及不同厚度的全锆贴面的平均断裂载荷值为:A组:(1275.9±205.6)N(0.5 mm厚度);(2320.6±296.7)N(1.0 mm厚度);(4909.1±399.3)N(1.5 mm厚度);B组:(1404.5±215.3)N(0.5 mm厚度);(2509.8±380.6)N(1.0 mm厚度);(4917.8±320.4)N(1.5 mm厚度);C组:(1279.6±248.4)N(0.5 mm厚度);(2346.6±260.2)N(1.0 mm厚度);(4575.3±458.7)N(1.5 mm厚度)。随着厚度的增加,全锆贴面的抗折性能也随之增加,同一牙体预备形式不同厚度的断裂载荷有统计学差异,不同牙体预备形式同一厚度的断裂载荷无统计学差异。0.5 mm与1.0 mm组别的失效模式(修复体折裂)相似,与1.5 mm组别的失效模式(修复体与代型同时折裂)存在差异。结论:全锆贴面厚度对其抗折性能有影响,抗折性能随着其厚度的增加而增大。牙体预备形式对贴面的抗折性能没有影响。厚度为0.5 mm的全锆贴面可以满足临床需要。

Abstract: Objective:This study aimed to evaluate the effect of the thickness and tooth preparation on fracture strength and failure mode of full-contour zirconia occlusal veneer. Methods:Standard artificial molars of 3 different groups(n=72) were prepared:Group A: the occlusal surface was ground of evenly by 0.5 mm; Group B: the occlusal surface was ground of evenly by 0.5 mm, and the axial surface was parceled for 1.5 mm; Group C: the occlusal surface was ground of evenly by 0.5 mm, and cavity fixation was prepared. Three different models of full contour zirconia occlusal veneers were designed with different thickness of 0.5 mm, 1.0 mm, and 1.5 mm respectively using CAD/CAM system and Dental Designer software. Fracture strength and failure mode of zirconia occlusal veneers were measured and recorded. Results:The average fracture strength value of full-contour zirconia occlusal veneer with different thicknesses and different preparation were: Group A:(1275.9±205.6)N(0.5 mm),(2320.6±296.7)N(1.0 mm),(4909.1±399.3)N(1.5 mm);Group B:(1404.5±215.3)N(0.5 mm),(2509.8±380.6)N(1.0 mm),(4917.8±320.4)N(1.5 mm);Group C:(1279.6±248.4)N(0.5 mm),(2346.6±260.2)N(1.0 mm),(4575.3±458.7)N(1.5 mm). Fracture strength of full-contour zirconia occlusal veneers increased in parallel with the increase of their thickness. Fracture strength of occlusal veneers with different thickness and preparation was significantly different, while that of occlusal veneers with the same thickness but different preparation showed no significant difference. The failure mode of 0.5 mm group was similar to that of 1.0 mm group but different from that of 1.5 mm group. Conclusions:Thickness is a positive correlation factor for the fracture strength of full-contour zirconia occlusal veneers. Preparation has no effect on the fracture strength of occlusal veneers. 0.5 mm is an acceptable thickness for zirconia occlusal veneers in clinical practice.